IZYSHAPE - Tracers settings
👓 ESSILOR TESS
Tracer uses only OMA format, no particular changes to be made.
SERIE connection
Check that tracer is on protocol Pc OmaRs232 or Pc OmaRs232+ and that parameters are the same between tracer and application.
Flow: HandShake: Rts
NoFlow: HandShake: None
Serial cable must be connected to RS232 plug.
IP CONNECTION
Tracer must appear alone in the application if it is on the same network as the computer and accessible via ping by it. If tracer does not appear, add it manually with its IP address.
👓 HUVITZ
Default settings are supported by tracers.
- Change the number of points (500 by default)
- Example: → change to 1000 points: Menu / Menu DCS / Data Point / 1000 points select by using the L ou R / B (Both) to validate
- In case of doubt, do not hesitate to contact the tracer supplier directly.
👓 WECO - TRACE 2
Below is protocols list compatible with iZyShape :
- RSA protocol: Configure protocol in OMA mode (unavailable)
- RSC protocol: Configure tracer protocol in Nidek mode (Bit Rate : 9600)
Change procedure (not tested):
After switching on tracer, press CTL and +/- keys simultaneously.
Then your screen should display tracer mode. You will have to choose RSC mode as screenshot below and press Enter key.
👓 WECO - TRACE 3
OMA protocol, it is necessary to configure the plot so it operates in slave mode.
- OMA Mode = hardware
- Upload device = TRC
- Bit rate: 9600
- Stopbits: 2
- Handshake: None
👓 NIDEK - LT910
Use Nidek protocol in tracer setup. Change tracer protocol according to the desired sending mode by right-clicking on tracer and then clicking on Property.
In about 90% of the cases, the Nidek protocol is used. However, it is possible that you are using an OMA protocol, so you need to configure this in your iZyShape tracer settings.
👓 HOYA - GT3000
Use Hoya2 protocol in probe setup. It is possible to use OMA protocol for this probe but the handling is complex.
Some special features in Series configuration:
Default settings may not be enough:
If when retrieving a shape, iZyShape displays the following error messages:
- Parity error detected: 👉try to change the par value from Even to None or vice versa and try again
- Frame error detected: 👉 try to change the value of StopBits from 2 to 1 or vice versa.
- If at the moment of recovering a shape the probe goes into error after a moment: 👉change the parity from Even to None or vice versa AND change the value of StopBits from 2 to 1 or vice versa
👓 HOYA - GT5000
Use Hoya3 protocol in probe setup. It is possible to change the protocol to OMA but you have to contact Hoya for that.
👓 ESSILOR KAPPA - PCCOMV2
We have encountered a case where the probe did not respond until DTR was activated in the advanced configuration options of Series probe. (HandShake on Xon)
Tests are relatively easy due to the fact that one can ask the last form probed as can be.