If problems are encountered during tracer's installation, it is advisable to activate debug mode at first and to keep files of plotting on disc.
IP CONNECTION
Tracer is supposed to appear all alone. In case it does not appear:
- Close any tracing software on all stations brought to use this tracer
- If this does not work and tracer is in static ip, you must create a new IP tracer by entering correct IP address
- Clicking on connect button should display connections such as tracer's connection is made
SERIE CONNECTION
👉VERIFY THAT COM PORT USED IS THE RIGHT ONE
...
- USB / Serie cable is easily noticed in sense that disconnecting it from computer will make it disappear from the list of devices
- Serie / Serie cable will not change display so it will be necessary to find out which port is actually used by the cable.
DEBUG MODE
👉ABSENT LOG FILE FOLLOWING A PALPAGE
...
It is necessary to review COM port connection or connection information which are erroneous. Protocol used can also be wrong.
👉LOG FILE PRESENT BUT NO OMA FILE PRESENT IN TRACE FILE
If log file is size enlarged at each tracing but no OMA file is created in tracer folder after it, there is a problem during tracer discussion.
This usually comes from parity / StopBits or in rare cases BaudRate.
👉PRESENT LOG FILE AND OMA FILE PRESENT OR NOT PRESENT BUT CAN NOT READ THE OMA FILE ON A WEBSHOP
...